Air and climate generator groups

are designed to be connected to test chambers in order to achieve the desired climatic conditions.

The air and climate generator sets are designed to be connected to test chambers to achieve the desired climate conditions (cold, heat and humidity)The interconnection between the two is made by flexible hoses for supply and suction air conduction. These hoses shall have the minimum path to avoid their energy load losses.

The maximum temperature ranges in standard models are -60ºC and can reach up to +180ºC, the volume of the equipment can be 300, 750 and 2000 litres, for other sizes, please consult us.

Generators sets are used in all the industrial sectors, as they meet testing standards for any product or material, but they are more common in the transport and electronics sectors.

At Ineltec, we design and manufacture customized equipment according to customer specifications, modifying or extending standard features.

Characteristics
From 150 L to 2000 L
adjustable from -60°C to +180°C
From 10% to 95 %
Standards

PV 2005: 2000-09. According to IEC -60068-3-5 ** Average no-load values

. . .

Exterior: Made of white-lacquered aluminium sheeting.

  • Mobility: 4 swivel castors with 2 braking devices or 4 height adjustable legs.
  • Hoses: They shall be installed according to access orifices (supply and suction), they shall be interconnected between the generator set and the test chamber.
  • Access holes: Side wall penetrations Ø, the number and size will depend on the customer's needs.

Interior: Made of stainless steel, lightweight and corrosion resistant. All hardware in stainless steel.

The touch screen PC has the PROCAM-WIN installed. It is easy and intuitive software which performs programming tasks, acquisition, records, control and analysis of the results. 

  1. Manual or automatic programming
  2. Possibility of programming date and time of the test
  3. It allows recording notes during the tests
  4. Different access levels
  5. Maximum 11 operators
  6. Display and recording of tests in a graphical or tabular table
  7. Export to Excel or similar
  8. The setting of minimum and maximum alarms for temperature and humidity for each section
  9. It includes recording screens, incidences and maintenance.
Subscribe to our Newsletter
and receive all our news
Sign up
You have to accept the terms of use..