Chambers to know the coefficient of thermal transmission

to study and measure the energy efficiency of air conditioning equipment and thermal insulation

Double environment chamber

Dual environmental calorimètric chambers are made from two climatic chambers with different environmental conditions which are used to test different materials of construction. 

They are used to study and measure the energy efficiency, insulation and heat transmission coefficient of thermal insulation. For example, on walls, doors, windows, etc.

At Ineltec, we design and manufacture customised equipment according to customer specifications, modifying or extending the standard features.

 

Characteristics
From -30ºC to +70ºC
Adjustable from 15 % to 98 %
According to customers specifications
Standards

EN-ISO 8990:1997, EN 1934:1998, EN-ISO 12567-1, EN 14511, EN 12412-2-4, EN 13420, EN 1121, EN 16147, PV 3545, EN 14240, EN 14511, EN 14509, ASHRAE 37-2009 , AHRI 210.240-2003 , EN 14825, EN 1946, EN 1026, etc.

. .

Design and fabrication of two climatic chambers for testing samples of different construction materials. The movement of the mobile camera is managed manually and its closure it's done by pneumatic cylinders.

Exterior: white lacquered aluminium sheet.

  • Access door: Double weatherstrip with flexible silicone gasket.
  • Base: Height-adjustable feet prefabricated in steel profile type HB

Interior:  made of AISI 304 stainless steel

  • Interior illumination: LED strip.
  • Access holes: shall be installed for electrical and mechanical connections between the chambers and the measuring elements

Insulation:

  1. Density: 41 Kg/m3
  2. Coefficient of thermal conductivity : 0,0136 Kcal/m.h.ºC
  3. Coefficient of thermal transmission: 0,163 Kcal/ m² hºC
  4. Completely waterproof and resistant to corrosive attack by acids and alkalis

The touch screen PC has the PROCAM-WIN installed. It is easy and intuitive software which performs programming tasks, acquisition, records, control and analysis of the results. 

  1. Manual or automatic programming
  2. Possibility of programming date and time of the test
  3. It allows recording notes during the tests
  4. Different access levels
  5. Maximum 11 operators
  6. Display and recording of tests in a graphical or tabular table
  7. Export to excel or similar
  8. Setting of minimum and maximum alarms for temperature and humidity for each section
  9. It includes recording screen, incidences and maintenance.
Subscribe to our Newsletter
and receive all our news
Sign up
You have to accept the terms of use..