It's used to cause an acid rain effect and to perform corrosive sulphuric atmosphere testing

Kesternich chamber

Kesternich corrosion chambers, CHKEST model, simulate a saturated humidity atmosphere and a constant temperature combined with sulphur dioxide gas (SO²) to produce an acid rain effect and to perform corrosive sulphuric atmosphere tests.

The standard temperature range is +5 °C and can be up to +70 °C, the volume of the equipment can be 300 and 1000 litres, for other sizes please consult us. The Kesternich corrosion chamber is mainly used for improvement studies in the prevention of corrosion in the automotive and building materials sectors.

In Ineltec, we design and manufacture tailor-made equipment according to specifications, modifying or extending standard features.

Characteristics
300 and 1000 L
ambient temperature +5°C and up to +70°C
100%
Standards

DIN EN ISO 6988 / ASTM G87 / BS 3900-F8:1993 / EN ISO 3131 / DIN EN ISO 6270-2 / DIN 50018

Documents
Kesternich Chamber
. .

ExteriorMade of polypropylene inert to corrosion

  • Access door: Flexible double silicone gasket, size according to chamber volume.
  • Sight glass:  A window is located in the door to view the interior of the chamber
  • Mobility: The machine has 4 height adjustable legs
  • Dosing device: The SO2 gas dosing device, made of transparent methacrylate with metric graduation, is located on the outside of the chamber.

Interior:  Made of polypropylene inert to corrosion

  • Inner supports for hanging the pieces to be tested 

Dimensions: (HxWxL)

Volume 300 L 1000 L
Inside Dimension 800 x 710 x 550 1140 x 1040 x 570
Outside Dimension 1000 x 1000 x 1000 1140 x 1500 x 1100

Approximately, as it depends on the temperature range, the outer dimension may be larger.

The touch screen PC has the PROCAM-WIN installed. It is easy and intuitive software which performs programming tasks, acquisition, records, control and analysis of the results. 

  1.  Manual or automatic programming
  2.  Possibility of programming date and time of the test
  3. It allows recording notes during the tests
  4. Different access levels
  5. Maximum 11 operators
  6. Display and recording of tests in a graphical or tabular table
  7.  Export to excel or similar
  8.  The setting of minimum and maximum alarms for temperature and humidity for each section
  9. It includes recording screens, incidences and maintenance.